Tmp- 2.0.16 Isolate Tmp; Tmp- 2.0.17 Isolate Venous Pressure - Fresenius Medical Care 2008K2 Troubleshooting Manual

Hemodialysis system
Hide thumbs Also See for 2008K2:
Table of Contents

Advertisement

TMP- 2.0.16 ISOLATE TMP

Turn Dialysate Flow OFF i.e. Flow on/off lamp blinks!
a)
b) Call debug screen 0. If (and ONLY if) ALL EIGHT (8) Balancing Chamber Valve 'dots' (#31 - #38) are
REMAINING WHITE i.e. Dialysate Flow is OFF see part d! NOTE: If the Balancing Chamber Valves
are NOT remaining white allow debug screen 1's FILACT = 0 then they will!
Leaving Flow OFF till instructed
c)

TMP- 2.0.17 ISOLATE VENOUS PRESSURE

This procedure creates Venous Pressure (VEN) to keep Valve #25 open for subsequent tests!
a) Figures below, attach a syringe with a piece of tubing attached that will fit SNUG to the Level
Detector's INNER P
b) Call debug screen 1. Push on the syringe plunger until VEN (middle column) is
between 400 and 450.
c) Clamp the syringe tubing to HOLD the pressure!
d) If VEN falls more than 3 mmHg in thirty (30) seconds there is a leak at the P
module.
e) Leaving the clamp in place to keep VEN more than 450 see procedure number
see procedure number
port.
ven
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
TMP- 2.0.17
(page 456).
port OR inside the
ven
TMP- 2.2.5
(page 457).
Page 456

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

2008k

Table of Contents