TMP- 2.3.2 POSITIVE PRESSURE TEST

a) CLAMP the external meter's tubing segment! The external meter is invalid during the next test!
b) SLOWLY PUSH on the syringe plunger until screen 0's
WARNING!
DO NOT exceed +325 mmHg!
c) Clamp the Four-Way Assemblies 'to syringe' tubing segment. Does
(Yes or No)?
Yes
PDial HOLDS +/- 25 mmHg per minute! See parts a AND b below:
IMPORTANT!
a)
b) Proceed to
No
PDial does NOT hold! See procedure number
TMP- 2.3.3 PDIAL DOES NOT HOLD / ISOLATE POSITIVE PRESSURE LEAK
IMPORTANT!
a)
Return the dialyzer lines to the shunt AND CLOSE THE DOOR!
b) A procedure, in a different Section, is performed next. NOTE this page and procedure number
(TMP- 2.3.3) because you may prompted to return here.
c) BEFORE continuing to part d, proceed to
PRESSURE
TESTS.
d) If a leak was not located in part c, check the Four Way for a positive pressure leak!
e) Place the machine in HEAT DISINFECT and allow Temperature to increase to more than 70° C!
f)
CAREFULLY check the hydraulic compartment, the ENTIRE LENGTH of the dialyzer lines, and the
®
DiaSafe
housing for external leaks.
Return the dialyzer lines to the shunt and CLOSE THE DOOR!
page 461
, procedure number
page 496
LEFT BLANK INTENTIONALLY
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
PDial
= between positive (+)300 and (+)325!
PDial
HOLD +/- 25 per minute
TMP-
2.3.4.
TMP- 2.3.3
(page 460).
, to perform
SECTION 12 - INDUCED POSITIVE