Front Panel And Display Assembly; Remove Front-Panel Assembly - Fluke 6270A Service Manual

Pressure controller/calibrator
Hide thumbs Also See for 6270A:
Table of Contents

Advertisement

6270A
Service Manual

Front Panel and Display Assembly

To prevent injury and electrical shock, disconnect the mains
power cable prior to the removal procedure, shut off the supply
pressure and vent the system before removing any manifold,
valve or associated parts.

Remove Front-Panel Assembly

To remove the front-panel assembly, see Figure 28:
1. Shut off the supply pressure and vent the system.
2. Remove all pressure lines and fittings from the manifold.
3. Open the front panel latch .
4. Disconnect the two cables .
5. Use #2 Phillips to remove the three screws that secure the display assembly
to the front-panel support .
6. Pull the front panel from the front-panel support .
To install the front-panel assembly, reverse the procedure.