PM200 Calibration Definition

PM200 Range
Pressure
---------
G20M
0
G14M
0
G10M
0
G7M
0
G4M
0
G3.5M
0
G2.5M
0
G2M
0
G1.4M
0
G1M
0
G700K
0
G400K
0
BG250K
0
BG250K
2
-100
BG200K
0
2
BG200K
-100
A200K
2
BG100K
0
2
BG100K
-100
A100K
2
BG60K
-60
BG40K
-40
BG35K
-35
BG2.5K
-2.5
1.
A reference may be substituted if the recommended reference is unavailable. The substituted reference must be able to cover the MIN and MAX
pressures required for the calibration of the applicable PMM. The references given here are the preferred from an overall uncertainty and
calibration interval viewpoint. Other types of references, such as a Fluke Calibration 7250 Pressure Controller/Calibrator, can potentially be used
when the standard is maintained on a shorter calibration interval.
2.
Negative Gauge Calibration. Use PG Differential Mode of reference is PG7601/PG7607. Use 2465 Negative Gauge Mode if reference is 2465.
-100 kPa is the nominal minimum pressure. Depending on ambient laboratory pressure, the laboratory may not be able to achieve -100 kPa.
88
Table 7. PM200 Calibration Definition
MIN
MAX
Reference
Pressure
System
[kPa]
[kPa]
-------------
20 000
PG7202
14 000
PG7202
10 000
PG7102
7000
PG7102
4000
PG7102
3500
PG7601
2500
PG7601
2000
PG7601
1400
PG7601
1000
PG7601
700
PG7601
400
PG7601
250
PG7601
0
PG7607, 2465
200
PG7601
0
PG7607, 2465
200
2465
PG7607,
100
PG7601, 2465
0
PG7607, 2465
100
2465
PG7607,
60
PG7601
PG7607,
40
PG7601
PG7607,
35
PG7601
2.5
FPG8601
1
Piston-Cylinder
Reference Mode
1
-------------
200 kPa,
500 kPa,
1 MPa,
2 MPa /kg P-C
200 kPa,
500 kPa,
1 MPa,
2 MPa /kg P-C
100 kPa/kg P-C
100 kPa/kg P-C
100 kPa/kg P-C
100 kPa/kg P-C
100 kPa/kg P-C
100 kPa/kg P-C
50 kPa/kg P-C
50 kPa/kg P-C
20 kPa/kg P-C
20 kPa/kg P-C
10 kPa/kg P-C
5 kPa/kg P-C, (TL)
P-C
10 kPa/kg P-C
5 kPa/kg P-C, (TL)
P-C
(TL) P-C
5 kPa/kg P-C, 10
kPa/kg P-C, (TL) P-
C
5 kPa/kg P-C, (TL)
P-C
(TL) P-C
5 kPa/kg P-C, 10
kPa/kg P-C
5 kPa/kg P-C, 10
kPa/kg P-C
5 kPa/kg P-C, 10
kPa/kg P-C
FPG8601
Dwell Time
-------------
[s]
Gauge
60
Gauge
60
Gauge
60
Gauge
60
Gauge
60
Gauge
60
Gauge
60
Gauge
60
Gauge
60
Gauge
60
Gauge
60
Gauge
60
Gauge
60
Differential,
60
Negative Gauge
Gauge
60
Differential,
60
Negative Gauge
Absolute by
60
Vacuum
Gauge
60
Differential,
60
Negative Gauge
Absolute by
60
Vacuum
Differential
60
Differential
60
Differential
60
Gauge
90
Averaging
Time
[s]
30
30
30
30
30
30
30
30
30
30
30
30
30
30
30
30
30
30
30
30
30
30
30
30

Hide quick links: