Remove and Replace Handles

To remove or replace the Product handles, see Figure 11:
1. Use a 2.5 mm hex wrench to remove the three M4 x 8 socket cap screws
on the side of each handle .
2. With the same wrench, remove the two socket-cap screws on the front of
each handle .
1
2
Figure 11. Replace the Handles
Pressure Controller/Calibrator
Repair and Replacement
hxn201.eps
27