Fluke 6270A Service Manual page 107

Pressure controller/calibrator
Hide thumbs Also See for 6270A:
Table of Contents

Advertisement

Calibration Segment
Ascending Points
Descending Points
1. UUT is zeroed before setting pressure.
Calibration Segment
Ascending Points
Descending Points
1. UUT is zeroed before setting pressure.
Table 30. PM200-G10M Pressure Point Matrix
As Found / As Left Points
Pressure No.
1
0
2
2000
3
4000
4
6000
5
8000
6
10 000
7
8000
8
6000
9
4000
10
2000
11
0
Table 31. PM200-G14M Pressure Point Matrix
As Found / As Left Points
Pressure No.
1
0
2
2800
3
5600
4
8400
5
11 200
6
14 000
7
11 200
8
8400
9
5600
10
2800
11
0
Pressure Controller/Calibrator
Set Pressure
Dwell Time [s]
[kPa]
1
60
60
60
60
60
60
60
60
60
60
60
Set Pressure
Dwell Time [s]
[kPa]
1
60
60
60
60
60
60
60
60
60
60
60
Calibration Tables
Averaging Time
[s]
30
30
30
30
30
30
30
30
30
30
30
Averaging Time
[s]
30
30
30
30
30
30
30
30
30
30
30
101