Clean The Exterior; Pressure Leakage Test (On-Board Diagnostics) - Fluke 6270A Service Manual

Pressure controller/calibrator
Hide thumbs Also See for 6270A:
Table of Contents

Advertisement

Clean the Exterior

To clean the Product, wipe it with a cloth that is lightly dampened with water or
mild detergent. To clean the display, use a soft cloth lightly dampened with
alcohol.
Do not use aromatic hydrocarbons or chlorinated solvents for
cleaning. They can damage the plastic materials used in the
Product.

Pressure Leakage Test (on-board diagnostics)

1. From the main menu, touch Tasks or SETUP>Tasks.
2. Touch Leak Test.
3. Complete these sections:
a. Setpoint – Pressure set point on which to perform the leak test.
b. Stability – After the set point is achieved, the time in seconds to wait for
stability before the duration of the test begins.
c. Duration – The time in seconds to collect data for leak-rate calculation.
d. Mode – Defaults to external, not selectable.
e. Run – Runs the test.
f.
Exit – Exit menu.
With the test port plugged and no volume, it is expected to achieve
better than 0.04 %FS/Min rate.
Caution
Note
Pressure Controller/Calibrator
Maintenance
21