Calibration Segment
Ascending Points
Descending Points
Calibration Segment
Ascending Points
Descending Points
1. ATM: atmospheric pressure from laboratory barometer via PG system.
Table 42. PM600- A3.5M Pressure Point Matrix
As Found / As Left Points
Pressure No.
1
70
2
875
3
1750
4
2625
5
3500
6
2625
7
1750
8
875
9
70
Table 43. PM600- A7M Pressure Point Matrix
As Found / As Left Points
Pressure No.
1
ATM
2
1750
3
3500
4
5250
5
7000
6
5250
7
3500
8
1750
9
ATM
Pressure Controller/Calibrator
Set Pressure
Dwell Time [s]
[kPa]
60
60
60
60
60
60
60
60
60
Set Pressure
Dwell Time [s]
[kPa]
1
60
60
60
60
60
60
60
60
1
60
Calibration Tables
Averaging Time
[s]
30
30
30
30
30
30
30
30
30
Averaging Time
[s]
30
30
30
30
30
30
30
30
30
107