Siemens SIMATIC S5-90U System Manual page 453

Table of Contents

Advertisement

S5-90U/S5-95U
Positions 1 and 2 on the operating mode switch have no function.
If you set the switch to "no linearization", module 464-8MF21 functions just like module 464-8MF11.
The characteristic linearization is possible for the following temperature ranges:
PT 100:
-200° C (-328° F) to+850° C (1562° F)
Analog Value Representation
Table 15-29. Analog Input Module 464-8MF21, 2×PT 100 "with Linearization",
Resist-
Units
ance in
>1766
>400
1766
1702
1700
390.26
1400
345.13
1000
280.90
600
212.02
300
157.31
200
138.50
2
100.39
0
100.00
- 40
92.16
- 80
84.27
- 200
60.25
- 202
- 494
<- 494
*
In the overrange the current slope of the characteristic curve is maintained when leaving the linearized nominal range.
Table 15-30. Analog Input Module 464-8MF21, 2 x PT 100 "without Linearization" (Unipolar)
Resistance
Units
in
>4095
400,0
4095
399,90
2049
200,098
2048
200,0
1024
100,0
1
0,098
0
0,0*
*
Because of tolerances of components used in the module, the converted value can also be negative (e.g. FFF8
Unit: -1).
EWA 4NEB 812 6115-02b
to IEC 751
Tempe-
rature
High Byte
in °C
>883
0 0 1 1 0 1 1 1 0 0 1 1 0 0 0 1
883
0
0
1
1
0
851
0
0
1
1
0
850
0 0 1 1 0 1 0 1 0 0 1 0 0 0 0 0
700
0 0 1 0 1 0 1 1 1 1 0 0 0 0 0 0
500
0 0 0 1 1 1 1 1 0 1 0 0 0 0 0 0
300
0 0 0 1 0 0 1 0 1 1 0 0 0 0 0 0
150
0 0 0 0 1 0 0 1 0 1 1 0 0 0 0 0
100
0 0 0 0 0 1 1 0 0 1 0 0 0 0 0 0
1
0 0 0 0 0 0 0 0 0 0 0 1 0 0 0 0
0
0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0
- 20
1 1 1 1 1 1 1 0 1 1 0 0 0 0 0 0
- 40
1 1 1 1 1 1 0 1 1 0 0 0 0 0 0 0
- 100
1 1 1 1 1 0 0 1 1 1 0 0 0 0 0 0
- 101
1
1
1
1
1
- 247
1
1
1
1
0
<- 247
1 1 1 1 0 0 0 0 1 0 0 1 0 0 0 1
High Byte
0 1 1 1 1 1 1 1 1 1 1 1 1 0 0 1
0
1
1
1
1
1
1
0
1
0
0
0
0
0
0 1 0 0 0 0 0 0 0 0 0 0 0 0 0 0
0 0 1 0 0 0 0 0 0 0 0 0 0 0 0 0
0 0 0 0 0 0 0 0 0 0 0 0 1 0 0 0
0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0
(in steps of 0.5° C (0.9° F))
Low Byte
1
1
1
0
0
1
1
0
0
0
1
0
1
0
0
1
1
0
0
0
0
0
1
1
0
1
1
0
0
0
0
0
0
1
0
0
1
0
0
0
Low Byte
1
1
1
1
1
1
0
0
0
0
0
0
0
0
1
0
0
0
Module Spectrum
Range
Overflow
1
Overrange*
1
Nominal range
1
Overrange*
1
Overflow
Range
Overflow
Overrange
Nominal range
H
15-87