oxyCRG Display
The oxyCRG (oxygen CardioRespiroGram) display provides a
simultaneous presentation of the following three waveforms overlayed on
the Main Screen:
Approximately the last six minutes of information is displayed for each
parameter.
The oxyCRG display can help you to make early detections of respiratory
regulation problems in neonates, by comparing trend patterns, or finding
interrelationships among the three parameters.
•
Beat-to-beat heart-rate trend
•
An oxygen parameter trend
•
Compressed respiration wave
Main Screen Display with oxyCRG
oxyCRG Display
Setting up your Monitor 3-25