The Rear
Panel of the
M1109A
External
Alarm
Device
Controls and
Connectors of
the M1109A
External Alarm
Device
Notes—
M1109A External Alarm Device
4
2
3
1. Standby Switch
2. Human Interface Link Connector Out; this is a Philips HIL
connector used to output the information from the handheld keypad
to the Utility CPU function card in the computer module.
3. Human Interface Link Connector In; this is a Philips HIL connector
used to connect the handheld keypad.
4. Power on/off cable; this connects to the XGA Display Controller
Board C'DSPC'-XGA in the M1046B Computer Module.
The connector cover must be put back on after all cables have been
connected
The External Alarm Device should be mounted for maximum
visibility of the alarm indicators on the front. The device should not
be mounted directly against a wall in order to ensure that the alarm
Controls and Connectors
Monitor Installation and Patient Safety 11-29