Changing the Patient Category
These safety values effect the measurement range of each category. The
Adult category has the greatest range (systolic 30 to 270 mmHg) and the
Neonate has the smallest (systolic 30 to 130 mmHg). Therefore you
cannot use Neonate or Pediatric if the expected pressure values
are outside the measurement range for that category.
The following tables list the measurement ranges and safety values for
each patient category.
NBP Measurement Range (in mmHg) for Differing Patient
Categories
Measurement Range
Systolic
Mean
Diastolic
Cuff Size
Default Inflation Pressure 165 mmHg
NBP Safety Values for Differing Patient Categories
Safety Value
Continuous Pressure
Time to inflate
from 0 to 60 mm Hg
Max. inflation time
3-48
Setting up your Monitor
Patient Category
Adult
Pedi
30 to 270
30 to 180
20 to 255
20 to 160
10 to 245
10 to 150
A to E
Any Cuff
125 mmHg
Patient Category
Adult
Pedi
15 mmHg
5 mmHg
(for 120s)
(for 120s)
40s
20s
60s
40s
Neo
30 to 130
20 to 120
10 to 100
1 to 4 & A
100 mmHg
Neo
5 mmHg (for
60s)
20s
30s