Emc; Performance Requirements - Philips V24CT User's Reference Manual

Cms patient monitoring system volume 1
Table of Contents

Advertisement

EMC

Perfor-
mance
Require-
ments
Specification
Resolution (pixel addressability)
Vertical Refresh Rate
Red, Green and Blue Video Inputs
Vertical & Horizontal Multi-Sync Inputs
Video-Cable Connector
1. The display must either fulfill the requirements of the EN60601-1-2
{IEC 601-1-2}, or
it must conform to the requirements of the EMC standard for ITE
devices, EN50082-1: 1997 and EN50081-1 / CISPR 22.
2. The video cable between M1167A/77A CMS Patient Monitoring
System and the ITE display must not exceed a length of 3.0m. For
cables exceeding 3.0m EMC testing according to IEC 801-4:1988 has
to be done.
Philips cannot assure compliance with the ANSI/AAMI EC-13 Standard for
Cardiac Monitors, Heart-rate Meters and Alarms when using ITE displays.
Compliance with the ECG aspect ration and 25 mm/s 10% sweep-speed
specifications can only be assured when using the Philips Displays
M1092A, M1094B and M1095A. In addition, the following display
specifications are recommended:
Requirement or Value
1024 x 768
60
~0.7
5
HD15male
The CMS and V24 and V26 Patient Monitors 1-39
Using an ITE Display
Units
Dots x lines
Hz (non-interlaced)
V p-p
V TTL