Philips V24CT User's Reference Manual page 350

Cms patient monitoring system volume 1
Table of Contents

Advertisement

Controls and Connectors
The V24 and
The monitor has several connectors which are shown below.
V26
Connectors
Do not apply a voltage of more than +12V to any of the pins on the
defibrillator sync, parameter module rack, Philips monitoring network
(SDN) or equipotential ground connectors.
The secondary ground complies with IEC 601-1-1 as shown on page 10-4.
The connectors on the monitor are:
Note—
not daisy-chain the AC power through the AGM. This configuration may
degrade RFI performance.
11-34
Monitor Installation and Patient Safety
AC LINE INPUT
RS232 CONNECTOR
(Rear Housing with
Serial Prefix <3608)
PROTECTIVE
EARTH GROUND
RS232 CONNECTOR
(Rear Housing with
Serial Prefix >=3608)
1. AC Power Connector; this connector is used to connect the
monitor to an AC power source.
If you have a Philips M1026A Anesthetic Gas Module (AGM), do
2. Secondary Ground Screw; this is a posidrive screw to connect the
monitor chassis to earth ground per IEC 601-1-1 standard.
3. Equipotential Ground Post, this connects the monitor to the
equipotential ground system.
DEFIB
CONNECTOR
RACK
CONNECTOR
EQUIPOTENTIAL GROUNDING
PHILIPS MONITORING NETWORK (SDN)
SINGLE CONNECTOR
marge3.tif