Transferring data from a source monitor via a transit monitor to
a destination monitor requires special handling if:
– The source monitor and the destination monitor are:
— CMS Release G or later or
— V24 or V26 Release A.0 or later
and
– The transit monitor is:
— CMS Release E or F
— V24 Release E
— CTS
1. Plug the Data Transfer Module into the rack of the source
monitor.
2. Press the
Module
data to the module.
3. Plug the Data Transfer Module into the rack of the transit
monitor.
4. Select
Collect New Data
Refernce
to collect new data to the module without clearing the module.
5. Remove the module from the transit monitor and plug it into
the destination monitor.
6. Press the
Monitor
data to the monitor.
With specific configurations, data loss can occur when transferring
Note—
data to a V24 or V26 Release B.0 via the Data Transfer module.
hardkey on the module and transfer all
in the Data to Module task window
Pts
hardkey on the module and transfer the
Types of Transfer
Data Transfer 10-13