PAT700 Total Organic Carbon Analyzer – Bottle Mode
14)
Anatel
If the selection has been made to use the conductivity meter verification resistor,
install the resistor onto the analyzer's 9-pin printer/calibration port, as illustrated in
Figure
6-14.
• If the calibration is being performed on a schedule, the conductivity meter test
dialog box does not appear, and the calibration occurs automatically.
• Otherwise, the conductivity meter test dialog box appears when the analyzer
passes the conductivity meter test. See
Verification resistor
Figure 6-14 Verification resistor installed on printer/calibration port
Figure 6-15 Conductivity meter test dialog box
Figure
6-15.
Page 87
Operator Manual