Hach ANATEL PAT700 Operator's Manual page 90

On-line total organic carbon analyzer
Hide thumbs Also See for ANATEL PAT700:
Table of Contents

Advertisement

Page 80
5)
6)
7)
8)
9)
Operator Manual
To use default settings, touch the use default settings radio button in the TOC
calibration setup dialog box. To run any other configuration, touch the use custom
settings radio button, then define the desired configuration. (See
page
79.) The following procedure relies on the recommended default setting.
Touch the forward icon. The OASIS door will unlock for five seconds. If the door is
not opened within five seconds, the unlock icon can be used to unlock the door
manually.
Install the standards bottles following the instructions as shown in the load bottles
dialog box and ensuring that you place the standards bottles in the correct
positions. See
Figure
• Bottle 1 = blank
• Bottle 2 = 250 ppb sucrose
• Bottle 3 = 500 ppb sucrose
• Bottle 4 = 750 ppb sucrose
Figure 6-5 Load bottles dialog box, TOC calibration
Bottle positions graphics
Unlock icon
energizes the bottle bay door lock for five seconds, allowing you to
open the door.
Back icon
takes you to the run standards dialog box.
Forward icon
takes you to the review test setup dialog box if RFID tagged bottles
are being used (see
bottles that are not RFID tagged are being used (see
Cancel icon
returns you to the previous dialog box. All changes made in any of the
dialog boxes are lost.
Close the bottle bay door.
Touch the forward icon. If bottles have been properly loaded and the RFID tags can
be read, the review test setup dialog box appears. See
not have RFID tags, the enter bottle information dialog box appears. See
Figure
6-7.
Bottle Mode – PAT700 Total Organic Carbon Analyzer
6-5. From left:
show you the order in which to install the bottles.
Figure
6-6), or to the enter bottle information dialog box if
Figure 6-4,
Figure
6-7).
Figure
6-6. If the bottles do
Anatel