Hach ANATEL PAT700 Operator's Manual page 119

On-line total organic carbon analyzer
Hide thumbs Also See for ANATEL PAT700:
Table of Contents

Advertisement

PAT700 Total Organic Carbon Analyzer – Alarms
Table 7-3
Warning alarm messages
Code
Message
131
UnableToStart4to20mAOutputs
132
UnableToStartPumpTest
133
UnableToStopPumpTest
134
UnableToCalibratePump
135
UnableToExportDataLog
136
UnableToExportAuditTrail
137
UnableToAddUser
138
UnableToEditUser
139
UnableToDeleteUser
141
UnableToAcknowledgeAlarms
142
UnableToDeleteAlarms
143
UnableToStartSelfClean
144
UnableToStopSelfClean
145
UnableToStartOnlineManualSample
146
UnableToStopOnlineManualSample
147
UnableToRestoreFactoryDefaults
148
UnableToStartVoltageTest
149
UnableToStartPlumbingTest
150
UnableToPrintDataLog
151
LampExtinct
153
Bottle1MarkedAsUsed
154
Bottle2MarkedAsUsed
155
Bottle3MarkedAsUsed
156
Bottle4MarkedAsUsed
157
ExcursionBottle1NotEmpty
158
ExcursionBottle2NotEmpty
Anatel
Alarm condition
Analyzer cannot start 4-20 mA output test.
Analyzer cannot start pump test.
Analyzer cannot stop pump test.
Analyzer cannot calibrate pump.
Analyzer cannot export data log.
Analyzer cannot export audit trail.
Analyzer cannot add a user.
Analyzer cannot edit a user.
Analyzer cannot edit a user.
Analyzer cannot acknowledge alarm.
Analyzer cannot delete alarms.
Analyzer cannot start self clean.
Analyzer cannot stop self clean.
Analyzer cannot start online manual sample.
Analyzer cannot stop online manual sample.
Analyzer cannot restore factory defaults.
Analyzer cannot start voltage test.
Analyzer cannot start plumbing test.
Analyzer cannot print data log.
• Analyzer detects that current lamp has failed.
• Replace UV lamp.
• Bottle installed in location 1 is marked as used.
• Replace bottle with a valid standard.
• Bottle installed in location 2 is marked as used.
• Replace bottle with a valid standard.
• Bottle installed in location 3 is marked as used.
• Replace bottle with a valid standard.
• Bottle installed in location 4 is marked as used.
• Replace bottle with a valid standard.
• Excursion bottle 1 is not empty.
• Remove excursion bottle 1 and replace with empty bottle.
• Excursion bottle 2 is not empty.
• Remove excursion bottle 2 and replace with empty bottle.
Page 109