Hach ANATEL PAT700 Operator's Manual page 92

On-line total organic carbon analyzer
Hide thumbs Also See for ANATEL PAT700:
Table of Contents

Advertisement

Page 82
10)
Operator Manual
C of A value text box
standard that is being used.
Volume text box
allows you to enter the volume of water, in mL, for the specified
standards bottle.
Expiration date text box
yyyy-mm-dd format, and time, in hh:mm:ss format, for the specified TOC calibration
standard.
Forward icon
takes you to the review test setup dialog box. See
Cancel icon
returns you to the load bottles dialog box.
Figure 6-8 Schedule test dialog box
Date text box
allows you to enter the date the test will run. The default value is
tomorrow's date. You may enter any valid date, in ISO "yyyy-mm-dd" format, that is
within three days from the loading of the standards.
Time text box
allows you to enter the time the test will run. The default time is the
current time. You may enter any valid time in hh:mm:ss format.
Back icon
takes you to the previous dialog box. All changes made in this dialog box
are retained.
Finish icon
takes you back to the home screen. All changes made in any of the
dialog boxes are retained and the test is scheduled to run. Prior to exiting this dialog
box, the analyzer verifies that the scheduled date and time is within three days of
the current date and time. It also verifies that none of the bottles will expire prior to
the scheduled date.
When the test running dialog box appears, the analyzer begins running the test.
See
Figure
6-9.
Bottle Mode – PAT700 Total Organic Carbon Analyzer
allows you to enter the certified analysis value for the
allows you to enter an appropriate expiration date, in ISO
Figure
6-6.
Anatel

Advertisement

Table of Contents
loading

Table of Contents