Page 90
8)
9)
10)
11)
Operator Manual
Figure 6-17 Load bottles dialog box, TOC validation
Bottle positions graphics
Unlock icon
energizes the bottle bay door lock for five seconds, allowing you to
open the door.
Back icon
takes you to the previous dialog box.
Forward icon
takes you to the review test options dialog box.
Cancel icon
returns you to the run standards dialog box. All changes made in any
of the dialog boxes are lost.
Close the bottle bay door.
Touch the forward icon. If bottles have been properly loaded and the RFID tags can
be read, the review test setup dialog box appears. See
Otherwise, the enter bottle information dialog box appears. See
page
81.
When the test running dialog box appears, the analyzer begins running the test.
• After verifying the presence of the appropriate standards, the analyzer performs a
temperature stabilization. Once the temperature has stabilized, the blank and
standards bottles are analyzed in the proper sequence.
• Once all the bottles have been analyzed, the analyzer compares the current results
to the standard calibration. If the average of each standard is within ±15% of the
expected value, the validation is accepted. If any of the averaged values are greater
than ±15% of the expected value, the validation is rejected.
• The dialog box displays the current status of the test and lists the results of each
repetition and bottle as it is running.
When the validation is completed, the test summary dialog box appears, with the
results of the validation and a notification of whether the validation failed or passed.
The results also appear on the printout if a printer is attached. See
page
83. Touching the forward icon takes you to the run again dialog box
(Figure 6-12, page
85). If the test failed or you have rejected the calibration results,
at the run again dialog box, touch the checkmark icon to run the validation again or
the cancel icon to cancel the run.
Bottle Mode – PAT700 Total Organic Carbon Analyzer
show you the order in which to install the bottles.
Figure 6-6, page
81.
Figure 6-7,
Figure 6-10,