Pump Test - Hach ANATEL PAT700 Operator's Manual

On-line total organic carbon analyzer
Hide thumbs Also See for ANATEL PAT700:
Table of Contents

Advertisement

Page 120
6)

Pump test

The pump test icon allows you to test pumps. See
Before the pump test, you must place a 50 mL graduated cylinder at the analyzer's water
outlet. No tubing should be used. A bottle must be installed in position 2 to enable the pump
to be primed. When you touch the run icon, the analyzer pumps water for 30 seconds, then
prompts you to input the volume pumped, as measured in the graduated cylinder. After you've
input the volume, touch the return icon to accept the results or the cancel icon to reject them.
Calibrate pump icon
Start/stop pump icon
pump valve and the grab sample bottle valve, then starts the pump. The pump runs
indefinitely until you press the stop pump icon. Both valves then close, and the pump stops.
Prime pump icon
The pump runs long enough to prime, then it stops. Both valves then close.
Load bottles icon
Unload bottles icon
page
Return icon
Operator Manual
If a test is chosen immediately, the analyzer asks if you wish to install a bottle. If yes
is selected, the load bottles dialog box illustrates the appropriate port in which to
install the bottle. The enter bottle information screen requires you to enter the
bottle's volume. The next button then starts the test.
opens the calibrate pump dialog box. See
starts and stops the pump. Touching the start pump icon opens the
opens the pump valve and the grab sample valve, then starts the pump.
opens the load bottles dialog box. See
takes you to the unload/replace bottles dialog box. See
119.
returns you to the diagnostics dialog box.
Diagnostics – PAT700 Total Organic Carbon Analyzer
Figure
Figure 8-12 Pump test dialog box
8-12.
Figure
8-13.
Figure 8-10, page
119.
Figure 8-11,