Hach ANATEL PAT700 Operator's Manual page 41

On-line total organic carbon analyzer
Hide thumbs Also See for ANATEL PAT700:
Table of Contents

Advertisement

PAT700 Total Organic Carbon Analyzer – Installation
Example:
Read command:
Read format:
"HH:mm:SS
HH:mm:SS
Response:
"00:01:00
00:01:00
Write format:
"HR HH:mm:SS
HH:mm:SS
Write command:
"HR 00:01:00
1 1 Sensor 0 1 0 3 0 1 0 0 0 0 1 1
00:01:00
Get number of times UV lamp is turned on, hours left, voltage and status of both UV lamps
"HH"
Example:
Read command:
Response:
"UVLamp1:
"UVLamp2:
Set/get user-calibrated
"HC"
conductivity slope
Example:
Read command:
Response:
Write format:
Write command:
Set/get user TOC slope
"HT"
Example:
Read command:
Response:
Write format:
Write command:
Anatel
"HR"
HH:mm:SS
u u u u s u u u u u u u u u u u u
u u f u u u u u u f"
00:00:00
50 0 1 1 Sensor 0 1 0 3 0 1 0 0 0 0 1 1
25 00 1.000 000 020 005 095 0 0 1.000"
HH:mm:SS
u u u u s u u u u u u u u u u u u
u u f u u u u u u f"
00:00:00
50
25 00 1.000 000 020 005 095 0 0 1.000"
"HH"
4 4300
1.000000
0 4300
1.000000
Slope range should be between 0.9 and 1.1.
"HC"
"1.000"
"HC f"
"HC 0.95"
Slope range should be between 0.85 and 1.15.
"HT"
"1.000"
"HT f"
"HT 1.05"
0
Good"
Good"
Page 31