Hach ANATEL PAT700 Operator's Manual page 118

On-line total organic carbon analyzer
Hide thumbs Also See for ANATEL PAT700:
Table of Contents

Advertisement

Page 108
Table 7-3
Warning alarm messages
Code
Message
37
TEMP_LOW
38
TEMP_HIGH
39
RESIST_LOW
40
RESIST_HIGH
100
IncorrectBottleLoadedinPosition1
101
IncorrectBottleLoadedinPosition2
102
IncorrectBottleLoadedinPosition3
103
IncorrectBottleLoadedinPosition4
104
NotEnoughFluidinBottle1
105
NotEnoughFluidinBottle2
106
NotEnoughFluidinBottle3
107
NotEnoughFluidinBottle4
118
UnableToChangeLanguageSettings
119
UnableToReadLanguageSettings
120
UnableToChangeEthernetSettings
121
UnableToReadEthernetSettings
122
UnableToPrintSystemSettings
125
UnableToStartLampTest
126
UnableToStopLampTest
127
UvLampLifeExceeded
129
UnableToReadInformationFromBottle
130
UnableToWriteInformationToBottle
Operator Manual
Alarm condition
Temperature out of range low.
Temperature out of range high.
Resistivity out of range low.
Resistivity out of range high.
• Incorrect bottle detected in position 1.
• Analyzer cannot perform bottle test operation.
• Incorrect bottle detected in position 2.
• Analyzer cannot perform bottle test operation.
• Incorrect bottle detected in position 3.
• Analyzer cannot perform bottle test operation.
• Incorrect bottle detected in position 4.
• Analyzer cannot perform bottle test operation.
• Not enough fluid in bottle 1 to complete operation.
• Check bottle 1 before starting operation.
• Analyzer aborts operation and returns to its normal operating mode.
• Not enough fluid in bottle 2 to complete operation.
• Check bottle 2 before starting operation.
• Analyzer aborts operation and returns to its normal operating mode.
• Not enough fluid in bottle 3 to complete operation.
• Check bottle 3 before starting operation.
• Analyzer aborts operation and returns to its normal operating mode.
• Not enough fluid in bottle 4 to complete operation.
• Check bottle 4 before starting operation.
• Analyzer aborts operation and returns to its normal operating mode.
Analyzer cannot change language settings.
Analyzer cannot read language settings
Analyzer cannot change Ethernet settings.
Analyzer cannot read Ethernet settings.
Analyzer cannot print system settings.
Analyzer cannot start lamp test.
Analyzer cannot stop lamp test.
• Lamp hours of usage have been exceeded or UV Detect™ output
has dropped below acceptable limit.
• Lamp has been set to a status of marginal.
• Replace lamp.
Analyzer cannot read information from bottle.
Analyzer cannot write information to bottle.
Alarms – PAT700 Total Organic Carbon Analyzer