Hach ANATEL PAT700 Operator's Manual page 111

On-line total organic carbon analyzer
Hide thumbs Also See for ANATEL PAT700:
Table of Contents

Advertisement

PAT700 Total Organic Carbon Analyzer – Bottle Mode
2)
3)
Anatel
Cancel icon
returns you to the previous dialog box. All changes made in any of the
dialog boxes are lost.
Close the bottle bay door.
Touch the forward icon. The excursion mode setup dialog box appears. See
Figure
6-30.
Figure 6-30 Excursion mode setup dialog box
TOC limit text box
before an excursion sample is captured. You may specify a value from 1 to 2000
ppb. The default setting is 500 ppb.
Run now icon
immediately draws a water sample from the online source into the
excursion bottle and opens the bottle filling dialog box. See
bottle has been filled, the excursion captured dialog box appears. See
page
99.
Done icon
returns you to the home screen. When an excursion condition occurs,
the analyzer performs an excursion sample capture. Following the excursion, the
analyzer always returns to the home screen.
Cancel icon
takes you to the unload/replace bottles dialog box. All changes made
in any of the dialog boxes are lost.
allows you to enter a limit an online TOC analysis must reach
Page 101
Figure
6-27. Once the
Figure 6-28,
Operator Manual

Advertisement

Table of Contents
loading

Table of Contents