Hach ANATEL PAT700 Operator's Manual page 109

On-line total organic carbon analyzer
Hide thumbs Also See for ANATEL PAT700:
Table of Contents

Advertisement

PAT700 Total Organic Carbon Analyzer – Bottle Mode
4)
Anatel
Run now icon
immediately draws a water sample from the online source into the
excursion bottle and opens the bottle filling dialog box. See
Done icon
returns you to the home screen. When an excursion condition occurs,
the analyzer performs an excursion sample capture. Following the excursion, the
analyzer always returns to the home screen.
Cancel icon
returns you to the unload/replace bottles dialog box. All changes made
in any of the dialog boxes are lost.
When an excursion occurs, an excursion sample captured dialog box opens, and a
validation occurs. See
• Following the excursion and validation, a validation results dialog box opens that
contains the blank and 500 ppb sucrose data. The next arrow on the validation
results dialog box takes you to the validation summary dialog box.
• The analyzer writes the date, time, TOC value, analyzer serial number, status,
oxidation curve filename, conductivity, and temperature information to the RFID tag.
Figure 6-28 Excursion sample captured dialog box
Figure 6-27 Bottle filling dialog box
Figure
6-28.
Page 99
Figure
6-27.
Operator Manual

Advertisement

Table of Contents
loading

Table of Contents