Operator Manual
Figure 6-25 Excursion mode with validation, load bottles dialog box
Bottle positions graphics
Unlock icon
energizes the bottle bay door lock for five seconds, allowing you to
open the door.
Back icon
takes you to the excursion mode dialog box.
Forward icon
takes you to the excursion mode setup dialog box.
Cancel icon
returns you to the previous dialog box. All changes made in any of the
dialog boxes are lost.
Close the bottle bay door.
Touch the forward icon. The excursion mode setup dialog box appears. See
Figure
6-26.
Figure 6-26 Excursion mode setup dialog box
TOC limit text box
before an excursion sample is captured. You may specify a value from 1 to 2000
ppb. The default setting is 500 ppb.
Bottle Mode – PAT700 Total Organic Carbon Analyzer
show you the order in which to install the bottles.
allows you to enter a limit an online TOC analysis must reach