PAT700 Total Organic Carbon Analyzer – Diagnostics
Flow path selection drop down list box
and pump combinations. You may select none (default), bypass, flow through cell, flow
through cell with pump, sample from bottle 1, sample from bottle 2, sample from bottle 3,
sample from bottle 4, back flush through bottle 1, back flush through bottle 2, back flush
through bottle 3, or back flush through bottle 4.
Load bottles icon
Unload bottles icon
page
Anatel
None.
Selecting this option turns off the pump and closes all the valves.
Bypass.
Selecting this option opens only the bypass valve.
Flow through cell.
Selecting this option opens only the cell valve.
Flow through cell with pump.
valve, and turns on the pump.
Sample from bottle 1, bottle 2, bottle 3, or bottle 4.
selected bottle valve, the pump valve, and turns on the pump.
Back flush through bottle 1, bottle 2, bottle 3, or bottle 4.
opens the selected bottle valve and the cell valve.
Return icon
returns you to the diagnostics dialog box.
opens the select bottles to load dialog box. See
takes you to the unload/replace bottles dialog box. See
119.
Figure 8-8 Plumbing test dialog box
allows you to select from 12 pre-configured valve
Selecting this option opens the cell valve, the pump
Selecting this option opens the
Selecting this option
Figure
8-9.
Figure 8-11,
Operator Manual
Page 117