Operator Manual
Touch the << icon to open the sliding toolbar, then touch the bottles icon in the
toolbar, or touch the bottles icon in the lower left corner of the screen.
Touch the run standards button in the bottle mode dialog box. See
page
76.
Touch the conductivity calibration button in the run standards dialog box. See
Figure 6-2, page
77.
The OASIS door will unlock for five seconds. If the door is not opened within five
seconds, the unlock icon can be used to unlock the door manually.
Install the standards bottle following the instructions as shown in the load bottles
dialog box and ensuring that you place the standards bottle in the correct position.
See
Figure
6-13. From left:
• Bottle 1 = empty
• Bottle 2 = 100 µS potassium chloride conductivity test standard
• Bottle 3 = empty
• Bottle 4 = empty
Figure 6-13 Load bottles dialog box, conductivity calibration
Bottle positions graphics
Unlock icon
energizes the bottle bay door lock for five seconds, allowing you to
open the door.
Back icon
takes you to the run standards dialog box.
Forward icon
takes you to the review test options dialog box.
Cancel icon
returns you to the previous dialog box. All changes made in any of the
dialog boxes are lost.
Close the bottle bay door.
Touch the forward icon. If bottles have been properly loaded and the RFID tags can
be read, the review test setup dialog box appears. See
Otherwise, the enter bottle information dialog box appears. See
page
81.
Bottle Mode – PAT700 Total Organic Carbon Analyzer
show you the order in which to install the bottles.
Figure 6-1,
Figure 6-6, page
81.