Shock and Vibration
Random vibration profiles were developed in accordance with MIL-T-28800E
Note
and MIL-STD-810E Method 514. Test levels exceed those recommended in
MIL-STD-810E for Category 1 (Basic Transportation, Figures 514.4-1 through 514.4-3).
Safety
Note
For UL and other safety certifications, refer to the product label or the
Product Certification
Electromagnetic Compatibility
For the standards applied to assess the EMC of this product, refer to the
Note
Product Certification
Note
For EMC compliance, operate this device with shielded cabling.
© National Instruments Corporation
Functional shock (PXI-6508)................. MIL-T-28800 E Class 3
Operational random vibration
(PXI-6508) ............................................. 5 to 500 Hz, 0.31 grms, 3 axes
Nonoperational random vibration
(PXI-6508) ............................................. 5 to 500 Hz, 2.5 grms, 3 axes
The PCI-DIO-96/PXI-6508/PCI-6503 meets the requirements of the
following standards of safety for electrical equipment for measurement,
control, and laboratory use:
•
IEC 61010-1, EN 61010-1
•
UL 61010-1, CSA 61010-1
section.
This product meets the requirements of the following EMC standards for
electrical equipment for measurement, control, and laboratory use:
•
EN 61326 (IEC 61326): Class A emissions; Basic immunity
•
EN 55011 (CISPR 11): Group 1, Class A emissions
•
AS/NZS CISPR 11: Group 1, Class A emissions
•
FCC 47 CFR Part 15B: Class A emissions
•
ICES-001: Class A emissions
section.
(per Section 4.5.5.4.1); half-sine
shock pulse, 11 ms duration,
30 g peak, 30 shocks per face
A-5
PCI-DIO-96/PXI-6508/PCI-6503 User Manual
Appendix A
Specifications
Online