AC Current (Non-Sinewave) Specifications

Squarewave
Ranges [1]
2.9 to 65.999 mA
66 to 659.999 mA
0.66 to 4.39999 A
4.4 to 22 A
[1] All waveforms are peak-to-peak output ranges.
[2] Uncertainty is stated in peak-to-peak. Amplitude is verified using an rms-responding DMM.

AC Current, Squarewave Characteristics (typical)

Range
I<4.4 A @ 400 Hz

AC Current, Trianglewave Characteristics (typical)

Linearity to 400 Hz
0.3% of p-p value, from 10% to 90% point
1-Year Absolute Uncertainty,
Frequency
+ (% of output + % of range) [2]
%Output
0.01 to 10 Hz
5.0%
10 to 45 Hz
0.25%
45 to 1 kHz
0.25%
1 to 10 kHz
5.0%
0.01 to 10 Hz
5.0%
10 to 45 Hz
0.25%
45 to 1 kHz
0.25%
1 to 10 kHz
5.0%
10 to 45 Hz
5.0%
45 to 1 kHz
0.5%
1 to 5 kHz
5.0%
45 to 500 Hz
2.0%
500 to 1 kHz
5.0%
Risetime
25 µs
40 µs to 1% of final value
tcal + 5°C,
%Range
0.5%
0.5%
0.25%
0.5%
0.5%
0.5%
0.5%
1.0%
1.0%
0.5%
1.0%
0.5%
1.0%
Settling Time
Aberrations
<1% of p-p value, with amplitude >50% of range
Multi-Product Calibrator
Additional Specifications
Maximum
Current
Resolution
Two digits, e.g., 50 mA
Six digits on each
range
Two digits
Six digits on each
range
Two digits
Six digits on each
range
Two digits on each
range
Six digits on each
range
Overshoot
<10% for loads <100 Ω