Messages
Target value not found
T
40
Target value not found
This error occurs when the target is not found during the marker search analysis after
specifying the target and executing the CALC:MARK:FUNC:EXEC and
CALC:FUNC:EXEC commands. This error occurs also when the bandwidth is not found
after executing the bandwidth marker command, CALC:MARK:BWID:DATA?
-124
Too many digits
The number of digits of the argument of the decimal numeric-value-data element exceeds
255 with the preceding 0 removed (see 7.7.2.4.1, IEEE488.2).
-223
Too much data
The block-, expression-, or character-string-type program data that has been received
conforms with the standard. But it exceeds the amount that can be processed under the
condition of the memory or conditions specific to memory-related devices. In this
instrument, this error occurs when the number of characters exceeds 254 in a
character-string parameter.
54
Transform, Gate not allowed
This message appears when you turn on the gating/conversion function of the time domain
feature, while frequency offset is enabled, measurement points are set to two, or sweep
type is set to other than linear sweep
Turn off the frequency offset (SENS:OFFS OFF command), set the measurement points to
three or more and the sweep type to linear sweep, and then turn on the gating/conversion
function.
-211
Trigger ignored
This instrument receives and detects a trigger command ("TRIG") or an external trigger
signal. But it is ignored due to the timing condition (This instrument is not in the wait-for-
trigger state, for example). Change the setup so that a trigger command or an external
trigger signal can be sent after the instrument has entered the wait-for- trigger state.
U
300
Unable to estimate adapter length
Adapter length can not be estimated at zero span. Change the frequency span other 0.
-113
Undefined header
A command not defined in this instrument, though not illegal in the syntactic structure, has
been received. For example, when a message ":DISP:WIND1:TABL:MEM ON" is sent to
a correct program message ":DISP:WIND1:TRAC1:MEM ON," the message sent is
received as an undefined command by this instrument. See the command reference and use
correct commands.
This error occurs also when a port not existing on this model is specified in a command
specifying a port number as an index. Such commands are
CALC:FSIM:SEND:DEEM:PORTn:xxxx, CALC:FSIM:SEND:PMC:PORTn:xxxx,
CALC:FSIM:SEND:ZCON:PORTn:Z0:R, and SENS:CORR:EXT:PORTn:TIME; they
include PORTn as a part.