Table 7-3
Post-Repair Procedures for Digital Assemblies
Replaced
Required Adjustments
Assembly or
Correction Constants (CC)
Part
PCA Front
No adjustment needed
Key (A52)
Handler I/O
No adjustment needed
Board (A25)
GPIB Board
No adjustment needed
USB Interface
No adjustment needed
Card
PCA Digital
No adjustment needed
Board (A53)
Removable
Calibration of the Touch Screen.
Hard Disk
Drive
Power Supply
No adjustment needed
Assembly
LCD TPANEL
Calibration of the Touch Screen
Assembly
Chapter 7
Post-Repair Procedures
Post-Repair Procedures
Verification
Perform "To Execute the diagnostics Test" on
page 69
Perform "To Check the Front Panel" on page 82
Perform "To Execute the diagnostics Test" on
page 69
Perform "To Check the GPIB" on page 84
Perform "To Check the USB Interface Card" on
page 84
Inspect the Booting Process
Perform "To Execute the diagnostics Test" on
page 69
Perform "To Check the Video Output" on page
84
Inspect the Booting Process
Inspect the Booting Process
Inspect the Booting Process
Perform "To Execute the diagnostics Test" on
page 69
Perform "To Check the LCD" on page 83