Agilent Technologies E5071C ENA Service Manual page 244

Network analyzers
Hide thumbs Also See for E5071C ENA:
Table of Contents

Advertisement

Post-Repair Procedures
Post-Repair Procedures
Post-Repair Procedures
Table 7-1 Post Repair Procedures lists the required procedures that must be performed
after the replacement of an assembly or CPU. These are the recommended minimum
procedures to ensure that the replacement is successfully completed.
It is recommended to confirm the performance executing with all of adjustment and
performance tests as final step.
Table 7-1
Post-Repair Procedures for Analog Assemblies for Up to 8.5 GHz Options
Replaced
Required Adjustments
Assembly or
Correction Constants (CC)
Part
PCI-DSP Card
Perform the following required
(A51)
adjustment.
- Write ID
- Save to Flash memory
Reference
Perform the following required
Oven Board
adjustment.
(A5)
- OCXO
244
Verification
Perform "To Execute the diagnostics Test" on
page 69
Perform the following performance test.
- AUX Input Test
- Frequency Accuracy Test
- RF Output Level Accuracy and Flatness Test
- RF Output Level Linearity Test
- Trace Noise Test
- Crosstalk Test
- System Dynamic Range Test
- Noise Floor Test
- Dynamic Accuracy Test
- Uncorrected System Performance Test
Perform "To Check the External Trigger Input"
on page 84
Perform "To Execute the diagnostics Test" on
page 69
Perform the following performance test.
- Frequency Accuracy Test