Table 7-2
Post-Repair Procedures for Analog Assemblies for Over 8.5 GHz Options
Replaced
Required Adjustments
Assembly or
Correction Constants (CC)
Part
Receiver
Perform the following required
Module (A7)
adjustments.
- Source Output Power
- Receiver IF Range
- Receiver Port Characteristics
- Receiver Absolute Gain
Front End
Perform the following required
Module (A9)
adjustments
- Source Output Power
- Receiver Port Characteristics
- Receiver Absolute Gain
Doubler
Perform the following required
Module (A6)
adjustments.
- Source Output Power
- Receiver Absolute Gain
Chapter 7
Post-Repair Procedures
Post-Repair Procedures
Verification
Perform "To Execute the diagnostics Test" on
page 69
Perform the following performance test.
- Trace Noise Test
- Crosstalk Test
- System Dynamic Range Test
- Noise Floor Test
- Dynamic accuracy test
- Uncorrected system performance
Perform "To Execute the diagnostics Test" on
page 69
Perform the following performance test.
- Uncorrected System Performance Test
Perform "To Execute the diagnostics Test" on
page 69
Perform the following performance test.
- Frequency Accuracy Test
- RF Output Level Accuracy and Flatness Test
- RF Output Level Linearity Test
- Trace Noise Test
- Crosstalk Test
- System Dynamic Range Test
- Noise Floor Test
- Dynamic Accuracy Test