Agilent Technologies E5071C ENA Service Manual page 304

Network analyzers
Hide thumbs Also See for E5071C ENA:
Table of Contents

Advertisement

Messages
Port 2 receiver overload
NOTE
When FOM function or an external signal source is used over 8.5 GHz frequency
measurements with E5071C-Over 8.5 GHz Options, there is a possibility that overload
detection and power trip occurs even if you measure the devices under the maximum input
level. In that case, turn OFF the overload & power trip detect function after ensuring the
value of input power level.
222
Port 2 receiver overload
The input to Test Port 2 exceeds the maximum input level. The measurement value is not
correct. When a DUT is an amplifier or the like, this error may occur, damaging the
receiver in the worst case. Should this error occur with a passive part used as the DUT or
with nothing connected to the test port, this instrument is faulty. Contact an Agilent
Technology sales office or the company from which you bought the instrument.
If this error occurs, the stimulus signal output is automatically turned to off.
NOTE
When FOM function or an external signal source is used over 8.5 GHz frequency
measurements, there is a possibility that overload detection and power trip occurs even if
you measure the devices under the maximum input level. In that case, turn OFF the
overload & power trip detect function after ensuring the value of input power level.
223
Port 3 receiver overload (for 4 Port Option only)
The input to Test Port 3 exceeds the maximum input level. The measurement value is not
correct. When a DUT is an amplifier or the like, this error may occur, damaging the
receiver in the worst case. Should this error occur with a passive part used as the DUT or
with nothing connected to the test port, this instrument is faulty. Contact an Agilent
Technology sales office or the company from which you bought the instrument.
If this error occurs, the stimulus signal output is automatically turned to off.
NOTE
When FOM function or an external signal source is used over 8.5 GHz frequency
measurements, there is a possibility that overload detection and power trip occurs even if
you measure the devices under the maximum input level. In that case, turn OFF the
overload & power trip detect function after ensuring the value of input power level.
224
Port 4 receiver overload (for 4 Port Option only)
The input to Test Port 4 exceeds the maximum input level. The measurement value is not
correct. When a DUT is an amplifier or the like, this error may occur, damaging the
receiver in the worst case. Should this error occur with a passive part used as the DUT or
with nothing connected to the test port, this instrument is faulty. Contact an Agilent
Technology sales office or the company from which you bought the instrument.
If this error occurs, the stimulus signal output is automatically turned to off.
NOTE
When FOM function or an external signal source is used over 8.5 GHz frequency
measurements, there is a possibility that overload detection and power trip occurs even if
you measure the devices under the maximum input level. In that case, turn OFF the
overload & power trip detect function after ensuring the value of input power level.
304