Figure B-3
Agilent Recovery Program(1/3)
Step 6. Recovery function is performed.
Step 7. A message box appears after the completion of recovery process. Press any key to
continue.
Figure B-4
Agilent Recovery Program(1/4)
Step 8. When the recovery is finished, E5071C restarts.
Step 9. After restart, the screen for initial registration appears. Execute initial registration. For
information on the execution procedure, refer to "Initial Registration Procedure (HDD
Revision: CN925 or above)" on page 279.
NOTE
If the touch screen does not work after System Recovery, refer to "Reconfiguring Touch
Panel Settings" on page 281.