Agilent Technologies E5071C ENA Service Manual page 248

Network analyzers
Hide thumbs Also See for E5071C ENA:
Table of Contents

Advertisement

Post-Repair Procedures
Post-Repair Procedures
Table 7-2
Post-Repair Procedures for Analog Assemblies for Over 8.5 GHz Options
Replaced
Required Adjustments
Assembly or
Correction Constants (CC)
Part
Reference
Perform the following required
Oven Board
adjustment.
(A5)
- OCXO
Analog Base
Perform the following required
Module (A0)
adjustment.
- AUX Input
- Frequency Reference
Synthesizer
Perform the following required
Module (A1)
adjustment.
- Synthesizer Gain
- Source Output Power
- Receiver Port Characteristics
- Receiver Absolute Gain
248
Verification
Perform "To Execute the diagnostics Test" on
page 69
Perform the following performance test.
- Frequency Accuracy Test
Perform "To Execute the diagnostics Test" on
page 69
Perform the following performance test.
- AUX Input Test
- Frequency Accuracy Test
Perform "To Execute the diagnostics Test" on
page 69
Perform the following performance test.
- Frequency Accuracy Test
- RF Output Level Accuracy and Flatness Test
- RF Output Level Linearity Test
- Trace Noise Test
- Crosstalk Test
- System Dynamic Range Test
- Noise Floor Test
- Dynamic Accuracy test
- Uncorrected System Performance Test