Glp Report - A&D EK-15KL Instruction Manual

Precision bench scale
Hide thumbs Also See for EK-15KL:
Table of Contents

Advertisement

10-2. GLP report

 To print the GLP report to an AD-8121B printer, select the function setting "info 1"
and "pUse 1" and use MODE 3 of the printer.
 To output the GLP report to a personal computer, select the function setting "info
2" and "pUse 0."
Calibration report
1. Perform calibration according to "7-3. Calibration
using a calibration weight."
end
2.
appears when the calibration is
complete.
glp
3.
appears and calibration report is
output.
end
4.
appears again. Remove the weight.
Press the
ON/OFF
OFF or press the RE-ZERO key.
AD-8121 format "info 1"
key to turn the power
Manufacturer
Serial number
ID number
Calibration
executed
Calibration weight
Column for
signature
:
Space, ASCII 20h
CR: Carriage return, ASCII 0Dh
LF:
Line feed, ASCII 0Ah
General format "info 2"
`~~~~~~~~~~A~&~D<CRLF>
MODEL~~~
Model
S/N~~~~6A6123456<CRLF>
ID~~~~~~~~ABCDEF<CRLF>
DATE<CRLF>
Date
<CRLF>
Time
TIME<CRLF>
<CRLF>
CALIBRATED(EXT.)<CRLF>
CAL.WEIGHT<CRLF>
~~~~~+15000.0~~
SIGNATURE<CRLF>
<CRLF>
<CRLF>
---------------~<CRLF>
<CRLF>
<CRLF>
31
To output the results
EK-15KL<CRLF>
<CRLF>
g

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ek-30kl