7.2 Models with internal weight (KERN ALJ/PLJ)

For non verified balances four adjustment possibilities are available in the menu.
Menu settings
In weighing mode press the MENU button and keep it pressed until
the acoustic signal gets mute. The first menu item „units" is displayed.
Press MENU button repeatedly until „calib" appears and confirm using
PRINT button.
Use the MENU key to choose between the following settings:
AUT-CAL
I-CAL
E-CAL
TEC-CAL
Take over selection using the PRINT button.
The balance returns to menu.
To finish the menu press the MENU button and keep it pressed until
the acoustic signal gets mute. The balance returns automatically into
weighing mode.
With activated AUT-CAL function the internal adjustment is
AUT-CAL
automatically started when the balance
Factory setting of
verified balances
The automatic adjustment function is always enabled. You can
start adjustment at any time by pressing the CAL-key manually.
26
Automatic adjustment with internal weight.
Factory settings of verified balances.
Adjustment with internal weight after having pressed
CAL, locked in verifiable units.
Adjustment with external weight
locked for verifiable units
not documented
• after the weighing balance was disconnected from
the mains
• after pressing ON/OFF in stand-by mode
• after a temperature change of 1.5 °C
with non loaded weighing plate / zero display
• after a time interval of 20 min
with non loaded weighing plate / zero display
ALJ/ALS/PLJ/PLS-BA-e-1135