Summary of Contents for Precision Medical EasyFlow5
Page 1
U S E R M A N U A L 4300 Model: PM Series SAVE THESE INSTRUCTIONS For the most current manual revision, please visit our Website: www.precisionmedical.com 300 Held Drive Tel: (+001) 610-262-6090 Northampton, PA 18067 USA Fax: (+001) 610-262-6080 ISO 13485 Certified...
DISPOSAL INSTRUCTIONS ..................7 REPLACEMENT PARTS ....................7 LIMITED WARRANTY ....................7 RECEIVING / INSPECTION Remove the Precision Medical, Inc. PM4300 Series Oxygen Concentrator from the packaging and inspect for damage. If there is any damage, DO NOT USE and contact your Provider. INTENDED USE The Precision Medical, Inc.
SAFETY INFORMATION - WARNINGS AND CAUTIONS Indicates an imminently hazardous situation which, DANGER if not avoided, will result in death or serious injury. Indicates a potentially hazardous situation which, if WARNING not avoided, could result in death or serious injury. Indicates a potentially hazardous situation which, CAUTION if not avoided, may result in minor or moderate...
Page 4
DANGER • DO NOT smoke, allow others to smoke or have open flames near the Concentrator when in use. • DO NOT use near any type of flame or flammable/explosive substances. • The Concentrator is not suitable for use in an enriched environment of flammable anesthetic mixture with air, oxygen or nitrous oxide.
SPECIFICATIONS For technical specifications call Precision Medical or visit www.precisionmedical.com. Operating Temperature: 50˚F to 95˚F (10˚C to 35˚C) Storage/ Transport Temperature: -29˚F to 140˚F (-34˚C to 60˚C) Operating / Storage / Transport Relative Humidity: Up to 95% Non-condensing Electrical (AC Power): 115 VAC ±10%, 60 Hz, 3.2 Amps;...
Fire Stop Check Valve to stop the flow of gas towards the patient in the case that the cannula becomes ignited. The Fire Stop Check Valve should be located as close to the patient as reasonably practicable. Fire Stop Check Valve - Precision Medical Part # 507706...
ALARM CONDITIONS Oxygen level is below 83%, continue to use but contact your provider. Oxygen level is too low; turn the concentrator “OFF”, connect to another source of Oxygen and contact your provider. Flow is lower than expected; clear blocked tubing or cannula. Flashing Power ON, but temperature is too high to start compressor;...
Compressor Inlet Filter 507919 LIMITED WARRANTY AND LIMITATION OF LIABILITY Precision Medical, Inc. warrants that the Oxygen Concentrator, (the Product), will be free of defects in workmanship and/or material for the following period: Oxygen Concentrator limited three (3) years from date of purchase Precision Medical, Inc.