Precision Medical Vac Trap PM9000 User Manual

Vacuum safety trap bottle

Advertisement

U S E R
Vacuum Safety Trap Bottle
CAUTION
300 Held Drive
Northampton, PA 18067 USA
ISO 13485 Certified
M A N U A L
Vac Trap
SAVE THESE INSTRUCTIONS
Federal (USA) law restricts this device to sale
by or on the order of a physician.
Model: PM9000
Tel: (+001) 610-262-6090
Fax: (+001) 610-262-6080
www.precisionmedical.com
(shown)

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Vac Trap PM9000 and is the answer not in the manual?

Questions and answers

Summary of Contents for Precision Medical Vac Trap PM9000

  • Page 1 U S E R M A N U A L Vac Trap Vacuum Safety Trap Bottle Model: PM9000 (shown) SAVE THESE INSTRUCTIONS Federal (USA) law restricts this device to sale CAUTION by or on the order of a physician. 300 Held Drive Tel: (+001) 610-262-6090 Northampton, PA 18067 USA Fax: (+001) 610-262-6080...
  • Page 2: Receiving Inspection

    RECEIVING / INSPECTION Remove the Precision Medical, Inc. Vac Trap from the packaging and inspect for damage. If there is any damage, DO NOT USE and contact your Provider. INTENDED USE To protect suction regulators and central vacuum systems from liquid contamination.
  • Page 3: Safety Information - Warnings And Cautions

    SAFETY INFORMATION - WARNINGS AND CAUTIONS Indicates a potentially hazardous WARNING situation which, if not avoided, may result in death or serious injury. CAUTION Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury. Used without the safety alert symbol, CAUTION indicates a potentially hazardous...
  • Page 4: Typical Installation Diagram

    TYPICAL INSTALLATION DIAGRAM CAUTION The Vac Trap must be in the vertical position to function. PARTS LIST Description Part # Float Assembly 503234 1731 Tubing Nipple 9073...
  • Page 5: Maintenance

    DO NOT USE if damaged. RETURNS Returned products require a Returned Goods Authorization (RGA) number, contact Precision Medical, Inc. All returns must be packaged in sealed containers to prevent damage. Precision Medical, Inc. will not be responsible for goods damaged in transit. Refer to Precision Medical, Inc.
  • Page 6 TROUBLESHOOTING Problem Probable Cause Remedy Float 1. Float neck dirty 1. Remove foreign Stuck substance 1. Leaking 1a. Check Suction connections connections & reconnect 1b. Tighten Jar 2. Turn Suction 2. Suction Source Source “ON” not turned “ON” 3. Empty / clean 3.
  • Page 7 Internal Cleaning Method: The vacuum trap can be cleaned using an enzymatic detergent solution and immersed during cleaning. It can be placed back into service using an ultrasonic cleaner, washer/disinfector, and/ or a hospital grade cleaner providing high level disinfection cleaners like Cidex OPA or the like. Caution should be taken not to exceed 200 degrees F when cleaning the Vac Trap, as temperatures higher than 200 degrees F will...
  • Page 8: Limited Warranty And Limitation Of Liability

    ORAL STATEMENTS DO NOT CONSTITUTE WARRANTIES. The representative of Precision Medical, Inc. or any retailers are not authorized to make oral warranties about the merchandise described in this contract, and any such statements shall not be relied upon and are not part of the contract for sale.
  • Page 9 Precision Medical, Inc. 300 Held Drive Northampton, PA 18067 USA Toll Free Tel: 800-272-7285 Toll Free Fax: 800-353-1240 Tel:(+001) 610-262-6090 Fax:(+001) 610-262-6080 www.precisionmedical.com ISO 13485 Certified 502056 Rev13 6/18 (?M) Printed in USA...