U S E R M A N U A L 4300 Model: PM Series (PM4351 Shown) SAVE THESE INSTRUCTIONS Federal (USA) law restricts this device to sale by or on the order of a physician. 300 Held Drive Tel: (+001) 610-262-6090 Northampton, PA 18067 USA Fax: (+001) 610-262-6080 ISO 13485 Certified...
RECEIVING / INSPECTION Remove the Precision Medical, Inc. PM4300 Series Oxygen Concentrator from the packaging and inspect for damage. If there is any damage, DO NOT USE and contact your Provider. INTENDED USE The Precision Medical, Inc. Stationary Oxygen Concentrator is intended to provide supplemental oxygen to persons requiring oxygen therapy.
SAFETY INFORMATION - WARNINGS AND CAUTIONS Indicates an imminently hazardous situation which, DANGER if not avoided, will result in death or serious injury. Indicates a potentially hazardous situation which, if WARNING not avoided, could result in death or serious injury. Indicates a potentially hazardous situation which, CAUTION if not avoided, may result in minor or moderate...
Page 5
DANGER • DO NOT smoke, allow others to smoke or have open flames near the Concentrator when in use. • DO NOT use near any type of flame or flammable/explosive substances. • DO NOT allow Concentrator to exceed “Operating or Storage Temperature Specifications”.
CAUTION DO NOT place any liquids on or near the Concentrator. If any liquid gets on or in the Concentrator, immediately turn “OFF” the power switch, unplug from the electrical outlet and contact your Provider . SPECIFICATIONS Dimensions: Length: 17.1 in (43.5 cm) Width: 12.3 in...
COMPONENT DESCRIPTION CAUTION Missing or illegible labels must be replaced, contact Precision Medical, Inc. Flow Setting Oxygen Flow Setting Oxygen Knob Knob Outlet Outlet ON / OFF ON / OFF Switch Switch Power Power Indicator Indicator Low Oxygen Action Concentration Required Alarm Indicator General...
OPERATING INSTRUCTIONS WARNING • Read this User Manual before operating the Concentrator. • Additional monitoring or attention may be required for patients using this Concentrator who are unable to hear or see alarms or communicate discomfort. CAUTION • Inspect the Concentrator for visual damage before use, DO NOT USE if damaged.
7. Place the cannula over your ears and position the prongs in your nose as instructed by your Provider or cannula manufacturer. 8. When the Concentrator is not in use, press the power On/Off ( / ) switch to “OFF” ( ) position.
Page 10
Visual Indicator/ Audible Probable Cause Remedy Problem Alarm Repeated Device overheating Move Concentrator to a cooler RED General Alarm Indicator is flashing beeping location. Ensure Concentrator while beeping, is positioned away from curtains device stops or drapes, hot air registers or functioning.
Returned products require a Returned Goods Authorization (RGA) number, contact Precision Medical, Inc. All returns must be packaged in sealed containers to prevent damage. Precision Medical, Inc. will not be responsible for goods damaged in transit. Refer to Precision Medical, Inc.
Oxygen Concentrator three (3) years from date of purchase Precision Medical, Inc. is NOT responsible for normal wear and tear, or any neglect or abuse of the product. Precision Medical, Inc. is NOT responsible for the shipping cost of the repair.