9.1 Operating modes (P8)

Under this menu item you can activate or test weigh the following operating modes of
the balance
P8 01: Parts count
P8 02: Checkweighing
P8 03: Filling
P8 04: Percent weighing
P8 05: Animal weighing
P8 06: Density
P8 07: Formulation
P8 08: Statistics:
Notice: Some operating modes may be combined (e.g. statistics/counting,
statistics/Percent determination). Further details are described in the
respective chapters.
Activation/deactivation of the operating modes:
In main menu call up menu item P8 "Operating mode" (see chapt. 8.1)
Press the
-key; the overview of the operating modes is displayed:
Use the arrow keys (see chapt. 8.1) to activate (1) or deactivate (0) the desired
operating modes.
Confirm your selection with the
Save (see chapter 8.1)
All activated operating modes are now stored in a sub-menu (M0 - M8) and can be
directly called up directly with the
44
20.02.05
Mode
P 8 01 Parts Count
02 Checkweighing
03 Filling
04 Perc nt
e
05 Animal
06 Density
07
Formulation
08
Statistics
key.
key:
20.02.05
Mode
M0 Basic weighing
M1 Parts counting
M2 Checkweighing
M3 Filling
M4 Percent
M5 Animal weighing
M6 Density
M7 Formulation
M8 Statistics
d is a b le d
1
e n a b le d
0
1
d is a b le d
d is a b le d
1
0
e n a b le d
d is a b le d
1
0
e n a b le d