9.1.6 Animal weighing (dynamic weighing)

For unsteady items to be weighed (e.g. animals) or strong vibrations you can activate
the dynamic weighing function with automatic or manual start. During a certain period
of time the balance determines weight values and calculates an average.
In automatic start measuring starts automatically as soon as the weight
changes.
In manual start you trigger measuring by actuating the
Operation
Call up menu item "M5 Animal weighing"
(see chap. 9.1)
Press
-button
01 Filter – filter for adjustment to the environmental conditions
Vibrations are filtered out by increasing the number of measuring cycles.
Use the cursor ( ► ) to select menu item "01 Filter"
Press the
-key; the current menu item blinks. Use the
arrow keys, to enter the sensitivity of the filter and
confirm by pressing the
very slow
slow
standard
fast
very fast
60
-key.
(insensitive,
very busy set-up location)
(sensitive
very quiet setup location)
-key.
Display