PHANTOM
2D
To check the accuracy of line laser level
To check the accuracy of line laser level (slope of plane)
Set up the instrument between two walls, the distance is 5 m.Turn on the Cross Line Laser and mark the point of cross laser line on
the wall.
Set up the instrument 0,5-0,7m away from the wall and make, as described above, the same masrks. If the difference {a1-b2} and
{b1-b2} is less then the value of "accuracy" (see spesifications), there is no need in calibration.
Example: when you check the accuracy of Cross Line Laser the difference is {a1-a2}=5 mm and {b1-b2}=7 mm. The
instrument's error: {b1-b2}-{a1-a2}=7-5=2 mm. Now you can compare this error with standard error.
If the accuracy of Cross Line Laser isn't corresponding with claimed accuracy, contact the authorized service center.
9
MEASUREMENT FOUNDATION
Need help?
Do you have a question about the Phantom 2D and is the answer not in the manual?
Questions and answers