9.1 Verified models

 Switch on device via ON-OFF and press the adjustment
switch at the same time.
 After that, during the balance's selftest, press
at the same time. „UnLoAd" appears.
 Acknowledge by
the weighing plate.
 The currently set adjustment weight will be displayed.
 To change by using the navigation buttons (see operating
instructions chap. 8) select the desired setting, the active
digit is flashing.
 Acknowledge by
 Carefully place adjusting weight in the centre of the
weighing plate. Wait for stability display, then press
„PASS" appears
 After the adjustment the balance will carry out a self-test.
Remove adjusting weight during selftest, the appliance will
return into weighing mode automatically.
An adjusting error or incorrect adjusting weight will be
indicated by the error message; repeat adjustment
procedure.
GAB-N-BA-e-1315