13.4 Error messages

Error message
Should other error messages occur, switch balance off and then on again. If the error
message remains inform manufacturer.
GAB-N-BA-e-1315
Description
Zeroing range exceeded
due to switching-on balance
0
or pressing
(normally
4% max)
Keyboard error
Value outside the A/D
changer range
Possible causes
Object on the weighing
plate
Overload when zeroing
Improper adjustment
Damaged weighing cell
Damaged electronics
Improper operation of the
balance
Damaged weighing cell
Damaged electronics