A "C" label near the calibration point indicates a custom point, while
an "H" indicates a HANNA standard value.
When the % D.O. range is calibrated, also the D.O. concentration range
is calibrated, and vice versa.
If no D.O. calibration has been performed or if calibration was cleared
using the "Restore factory calib." option the offset and slope values
are set to default, and the message "Factory calibration" is displayed.
Press <ESC> to return to the previous screen.
Conductivity
• From the "GLP" menu select the "Conductivity"
option.
• Data regarding the last conductivity calibration
will be displayed: calibration point, cell constant
value, calibration type (conductivity, absolute
conductivity or salinity), time and date of the cali-
bration.
• Use the arrow keys to scroll through the stored data for the last 5 calibrations.
Notes
A "C" letter near the conductivity calibration indicates a custom point,
while an "H" indicates a HANNA standard value.
If no conductivity calibration has been performed or if calibration was
cleared using the "Restore factory calib." option the offset and slope
values are set to default, and the message "Factory calibration" is dis-
played. Press <ESC> to return to the previous screen.
Turbidity
• From the "GLP" menu select the "Turbidity" option.
• Data regarding the last turbidity calibration
will be displayed: standards used, time and
date of the calibration.
• Use the arrow keys to scroll through the
stored data for the last 5 calibrations.
• Press <ESC> to return to the "GLP" menu.