Hanna Instruments HI 9813-5 Instruction Manual

Hanna Instruments HI 9813-5 Instruction Manual

Portable ph/ec/tds/°c meters
Hide thumbs Also See for HI 9813-5:

Advertisement

Instruction Manual
HI 9813-0
HI 9813-5 & HI 9813-6
Portable
pH/EC/TDS/°C Meters
w w w . h a n n a i n s t . c o m

Advertisement

Table of Contents
loading

Summary of Contents for Hanna Instruments HI 9813-5

  • Page 1 Instruction Manual HI 9813-0 HI 9813-5 & HI 9813-6 Portable pH/EC/TDS/°C Meters w w w . h a n n a i n s t . c o m...
  • Page 2: Table Of Contents

    If the repair is not covered by the warranty, you will be notified of Note: Save all packing material until you are sure that the instrument the charges incurred. If the instrument is to be returned to Hanna Instruments, first functions correctly. Any defective items must be returned in the obtain a Returned Goods Authorization number from the Customer Service department original packing together with the supplied accessories.
  • Page 3: Functional Description, Hi 9813-0 & Hi 9813-5

    FUNCTIONAL DESCRIPTION FUNCTIONAL DESCRIPTION HI 9813-6 HI 9813-0 AND HI 9813-5 1) 8-pin DIN connector for probe 1) 8-pin DIN connector for probe 2) Liquid Crystal Display 2) Liquid Crystal Display 3) pH range selection key 3) Green and Red LEDs for Cal-Check 4) mS/cm (EC) range selection key 4) mS/cm / ppm (EC/TDS) range selection key 5) EC/TDS calibration knob...
  • Page 4: Specifications

    SPECIFICATIONS OPERATIONAL GUIDE INITIAL PREPARATION Range 0.0 to 14.0 Each meter is supplied complete with a 9V battery. Remove the 0 to 1999 battery compartment cover on the back of the meter and install the mS/cm 0.00 to 4.00 battery while paying attention to its polarity. °C 0.0 to 60.0 (HI 9813-5 &...
  • Page 5: Ph Calibration

    • Select the appropriate measurement range (EC or TDS). pH CALIBRATION For greatest accuracy, frequent calibration of the instrument is recommended. The instrument should be recalibrated for pH: a) Whenever the electrode is replaced. b) At least once a month. •...
  • Page 6: Ph Values At Various Temperatures

    Note: The probe should be submerged approximately 4 cm (1½") into the EC/TDS CALIBRATION solution. The ChecktempC thermometer has to be located close to the probe. Note: If turning the knob the needed value can not be reached, Accessories needed: clean the probe (see the "Probe Maintenance"...
  • Page 7: Probe Maintenance

    4. if the red LED lights up again, calibration is needed (see "pH PROBE MAINTENANCE Calibration" section). PERIODIC MAINTENANCE Note: If it is not possible to calibrate, the probe has to be replaced Inspect the probe and the cable. The cable used for the connection to with a new one.
  • Page 8: Accessories

    HI 710016 Shockproof rubber boot, orange HI 710022 Blue protective case Hanna Instruments reserves the right to modify the design, construction and All rights are reserved. Reproduction in whole or in part is prohibited without the appearance of its products without advance notice.
  • Page 9 SALES AND TECHNICAL SERVICE CONTACTS Australia: Tel. (03) 9769.0666 • Fax (03) 9769.0699 China: Tel. (10) 88570068 • Fax (10) 88570060 Egypt: Tel. & Fax (02) 2758.683 Germany: Tel. (07851) 9129-0 • Fax (07851) 9129-99 Greece: Tel. (210) 823.5192 • Fax (210) 884.0210 Indonesia: Tel.