• The stability counter will count down and the message "Ready" and "Confirm"
will be displayed. Press <Confirm> to store the calibration point.
• After confirmation, the following messages are displayed: "Storing" and "Cali-
bration completed".
• Press <Measure> to return to the measurement screen.
• Press <OK> to return to the "Calibration" menu.
• To restore the factory calibration, select "Restore factory calib." in the "Pressure
calibration" menu and press <Select>.
51